2/3
B 60
Connection to the pumping circuit
G 0158 - Edition 07 - Januar
y 10
At the pump exhaust
Connection
Exhaust connection: DN 40 ISO-KF
At the pump inlet
Inlet isolation valve
(accessory)
This accessory avoids a sudden reverse flow of gas to the chamber
when the pump is stopped.
Connect the valve directly on the pump inlet flange using connecting
accessories.
Connect the electrical cable to the electronic cabinet inside the frame
(see
D 10
).
If the valve must be remote controlled, see
B 70
.
DN 50 ISO-KF for ADP 122.
DN 100 ISO-K for ADS 602, ADS 1202, ADS 1802.
Connection
Exhaust of corrosive, reactive, flammable, pyrophoric or oxidizing process
gases may result in severe injury or death.
Always connect the pump exhaust of the P and H versions to an exhaust
extraction system.
Ensure that all components in the exhaust pipeline have a maximum
pressure rating which is greater than the highest pressure that can be
generated in your system.
Make sure that any vacuum accessories connected to the inlet of the
pump can withstand a negative pressure of 1 bar with reference to the
atmospheric pressure.
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two