G 01063 - Edition 06 - December 06
3/4
B 130
Installation of exhaust heater accessory
Facilities at the rear
(exhaust in horizontal position)
Install the heating belts (
3
) on the
pipes, using the tightening cords.
Install the heat insulating jackets (
4
).
Exhaust heater
installation
Facilities on top (exhaust straight up)
The accessory is made up several heating belts to be installed on each
one of the exhaust pipe elements (straight up and bent).
10
13
17
Fit the pipe (
10
) on the pump exhaust and secure the
2 screws or the 2 nuts (
7
).
For facilities on top, fit the centering ring, the quick connect clamp, the
flowmeter panel and the banjo screw.
If the exhaust pipe is equipped with a pressure sensor, refit it.
Reassembly of the
equipped exhaust pipe
3
4
Install the heater devices on the
exhaust pipes.
Install the exhaust pipe in a reverse order then disassembly.
For ADS 1802P only
For ADS 1802P only
After the accessory assembly, perform an helium leak test on the pumping
line.