Saving and remote loading of pump
configuration (M4 monitoring)
At pump installation or replacement, it may be interesting to copy
the configuration (set of all the parameters and operation setting
programmed by the user) from a pump to another.
The pump configuration is memorized in the internal electronics
of M4 monitoring.
Enter in the menu by
SET
.
Go on using
+
to access
the MANAGEMENT menu
and valid by
ENTER
.
Go on using
+
to access
the CONFIG.SAVE menu
and valid by
ENTER
.
The pump configuration is then
duplicate into the hand-held remote
control memory.
Return to main menu and exit the setting mode by
SET
.
Disconnect the hand-held remote control.
Configuration saving
DEFINITION
SETTING
MAINTENANCE
>>> MANAGEMENT <<<
STORAGE
>>>> CONFIG SAVE <<<<
CONFIG. LOAD
OPERATING TIME
Loading and saving the parameters with the pump in operation may result
in a software conflict.
Perform these operations when the pump is switched on but with pumping
stopped.