G 01159 - Edition 05 - December 06
1/2
E 012
Draining of the water cooling circuit
Disconnect the water inlet and outlet quick connectors.
Connecting the
cooling circuit
IN
OUT
Facilities on the top
Facilities
at the rear
1/4 NPT
1/4 NPT
Facilities on the top
Facilities
at the rea
As the water circuit is equipped with water valve (normaly closed),
emptying operation must be done when the pump is running (so valves
are opened).
Prepare a flexible pipe (A) with a 1/4 inch quick male connector (to
be connected to compressed air circuit) and another flexible pipe (B)
with a 1/4 inch female connector.
Disconnect inlet and outlet water fittings on the pump.
Connect pipe (
B
) on the pump, at location ”
OUT
” and place the other
end of the pipe in a vessel (total water volume will be grater than
one litre).
Connect pipe (
A
) at location of ”
IN
” and connect the other end to a
compressed air circuit (2 to 5 bars).
A
air comprimé
récipient
B
vessel
compressed
air
adixen Vacuum Products - Operating instructions - ADP/ADS Series Two