G 01160 - Edition 03 - December 06
1/1
E 013
Saving of the pump configuration
Back up the settings of the pump to be replaced in order to enter them
into the new pump and ensure identical operations.
We advise not returning the hand-held remote control to the service
center.
The pump configuration is memorized in the internal electronics of M4
monitoring.
Enter in the menu by
SET
.
Go on using
+
to access
the MANAGEMENT menu
and valid by
ENTER
.
Go on using
+
to access
the CONFIG.SAVE menu
and valid by
ENTER
.
The pump configuration is then
duplicate into the hand-held remote
control memory.
Return to main menu and exit the setting mode by
SET
.
Disconnect the hand-held remote control.
Configuration saving
DEFINITION
SETTING
MAINTENANCE
>>> MANAGEMENT <<<
STORAGE
>>>> CONFIG SAVE <<<<
CONFIG. LOAD
OPERATING TIME
Loading and saving the parameters with the pump in operation may result
in a software conflict.
Perform these operations when the pump is switched on but with pumping
stopped.