Waveform Analysis/Search
11
Note
•
The interval over which GO/NO-GO determination is performed is set using Time Range1
and Time Range2 in the GO/NO-GO menu. Other items follow the settings specified in
the Measure menu.
•
If you are not going to perform GO/NO-GO determination using an external signal, make
sure to turn OFF Remote. Leaving it ON can cause instability in the start/stop operation of
data acquisition using the START/STOP key
Action: Set Action
The following five types of actions can be executed when the condition is met.
• Print the Screen Image Data: PRINT
Prints the screen image data on the printer (Printer (built-in printer), USB (USB
printer), or Net Print (network printer)) specified by Print to in the PRINT menu.
• Save the Screen Image Data: Image
Saves the screen image data to the destination (FD, Zip disk, PC card, internal HD
(optional), SCSI device, or USB storage device) specified in the IMAGE SAVE menu.
• Save the Waveform Data: Save to File
Saves the waveform data in binary, ASCII, or floating format to the destination (FD,
Zip disk, PC card, internal HD (optional), SCSI device, or USB storage device)
specified in the FILE menu. The save format is synchronized to data type (see
section 13.7) on the FILE menu.
• Buzzer
Sounds a buzzer.
• Send Mail
Sends an e-mail message to a specified address. (when the Ethernet interface option
is installed)
For the procedure in setting the address, see section 16.5, “Sending Periodic Mail or
Action Mail.”
Setting the Determination Range (Time Range1/Time Range2)
By default, the determination range is
±
5 divisions of the display frame on the time axis.
You can limit this range. The concept of the determination range is analogous to the
concept of the selectable range of cursor display position in cursor measurement. For
details, see section 11.5, “Selectable Range of Cursor Position.”
Executing/Aborting GO/NO-GO Determination
The operation starts when you press the Exec soft key. Waveform acquisition is
automatically stopped when the determination is complete. To forcibly stop the
operation, press START/STOP or the Abort soft key.
If Remote is turned ON, determination is executed using the input from the external
terminal. To forcibly stop the operation, press START/STOP.
Pressing the Exec (One Shot) soft key executes the same operation as the input from
the external terminal.
Save to File/PRINT/Image Operation
Operation is carried out according to the settings of FILE, PRINT, and IMAGE SAVE.
11.9 GO/NO-GO Determination Using Measured Waveform Parameters