19-1
IM 701210-06E
Specifications
19
Chapter 19
Specifications
19.1 Input Section
Item
Specifications
Number of input channels
16 ch 16-bit log (8 bits
×
2)
Type
Plug-in input unit
Number of slots
8 (2 channels per slot)
Maximum Record Length
Standard
2.5 MW/CH (16 analog ch 6 DSP ch 16-bit logic), 50 MW/CH max.
/M1 option 10 MW/CH (16 analog ch 6 DSP ch 16-bit logic), 250 MW/(1CH) max.
/M2 option 25 MW/CH (16 analog ch 6 DSP ch 16-bit logic), 500 MW/(1CH) max.
/M3 option 50 MW/CH (16 analog ch 6 DSP ch 16-bit logic), 1 GW/(1CH) max.
DSP channel (optional)
6 dedicated computation channels (DSP channels) provided internal to the DL750/DL750P.
Performs realtime computation of addition, subtraction, multiplication, and division (with or without
coefficients), filtering, differentiation, integration, and knocking filtering using the 16 analog input
channels as computation sources. (For detailed specifications of the computation function of the
DSP channels, see section 19.5, “Function.”)
Below are the characteristics of the DSP channels
• The allocation of the acquisition memory of DSP channels is the same as analog input channels.
• Can be used in all acquisition modes.
• Can be set as a trigger source of simple triggers, the OR trigger of enhanced triggers, and
window triggers.
• Computed waveform can be displayed in realtime even during roll mode display.
• Can be used as a target waveform of cursor measurements and automated measurement of
waveform parameters.
• Can be used as a target channel for the dual capture function.
19.2 Trigger Section
Item
Specifications
Trigger mode
Auto, auto-level, normal, single, single(N), log, and repeat (only during Chart Recorder mode on the
DL750P)
Trigger level range
CH1 to CH16:
±
10 div around 0
Trigger hysteresis
When observing voltage: Select
±
0.1 div,
±
0.5 div, or
±
1 div of the trigger level
When observing temperature: When observing temperature: Select
±
0.5
°
C,
±
1.0
°
C, or
±
2.0
°
C.
Trigger position
Can be set in 0.1% increments of the display record length
Trigger delay range
0 to 10 s (resolution is 100 ns)
Hold off time range
0 to 10 s (resolution is 100 ns)
Manual trigger key
Dedicated manual trigger key is available
Simple trigger
Trigger source
CH1 to CH16, EXT (signal input from the TRIG IN terminal), LINE (commercial power supply signal
that is connected), Logic A, Logic B, Time, and DSP1 to DSP6
Trigger slope
CH1 to CH16 and DSP1 to DSP6: Rising, falling, or rising/falling
EXT, LOGIC A, LOGIC B:
Rising or falling
Time trigger
Date (year/month/day), time (hour/minute), time interval (1 minute to 24 hours)
Enhanced trigger
Trigger source
CH1 to CH16, Logic A, and Logic B (AND and OR possible on each logic bit)
Trigger type
A -> B(N):
Trigger occurs nth time condition B becomes true after condition A becomes true.
Count:
1 to 255
Condition A:
Enter/Exit
Condition B:
Enter/Exit
A Delay B:
Trigger occurs first time condition B becomes true after specified delay following
condition A true.
Delay:
0 to 10 s (resolution is 100 ns)
Condition A:
Enter/Exit
Condition B:
Enter/Exit