19-41
IM 701210-06E
Specifications
19
19.13 Logic Probe Specifications
High-Speed Logic Probe (700986) Specifications
Item
Specifications
Number of input points
8
Input format
Non-isolated (the earth of each bit are common, the earth of the DL750/DL750P and the earth of
each bit are common)
Maximum input voltage
42 V (DC+ACpeak)(CAT I and CAT II, 30 Vrms), across probe tip and earth
(at a frequency of 1 kHz or less)
Response time
1
µ
s or less
Input impedance
100 k
Ω
or more
Threshold level
approx. 1.4 V
Isolation Logic Probe (700987) Specifications
Item
Specifications
Number of input points
8
Input format
Isolated (all bits are isolated)
Input connector
Safety terminal type (for banana plug)
×
8
Input switching
Can switch between AC/DC input for each bit
Input signal display
Can confirm H/L with the LED for each bit (lights when H)
Applicable input range
During DC input:
H/L detection of 10 VDC to 250 VDC
During AC input:
H/L detection of AC type of 80 VAC to 250 VAC 50/60 Hz
Threshold level
During DC input:
6 V
±
50% (Hi level: 10 to 250 VDC, Lo level: 0 to 3 VDC)
During AC input:
50 VAC
±
50% (Hi level: 80 to 250 VDC, Lo level: 0 to 20 VAC)
Response time
During DC input:
within 1 ms
During AC input:
within 20 ms
Input impedance
approx. 100 k
Ω
Maximum input voltage
250 Vrms
1
(CAT I and II)
(across H and L of each bit)
Maximum allowable
250 Vrms
1
(CAT I and II)
common mode voltage
(across input terminal H or L and earth)
Maximum allowable
250 Vrms
1
(CAT I and II)
voltage between bits
Withstand voltage
2000 VAC for 1 minute
(across input terminal and earth)
Isolation resistance
500 VDC, 10 M
Ω
or more
(across input terminal and earth)
Fuse
2
location
max. rated voltage
max. rated current
type
standard
H side of input terminal
250 V
50 mA
time lag
VDE/SEMKO
1.
Make sure the ACpeak voltage does not exceed 350 V and the DC voltage does not exceed 250 V when the frequency is 1
kHz or less.
2.
Because the fuses used by this instrument are all inside the case, they cannot be exchanged by the user.
If you believe the fuse inside the case is blown, please contact your nearest YOKOGAWA dealer.