18.2 Messages and Corrective Actions
Code
Message and Corrective Action
Section
887
A sampling speed was too fast, and it could be folded, and a real-time queue was needed.
7.7
Slow down a sampling speed.
888
A sampling speed was too fast, and it could be folded, and a real-time buffer was needed.
7.7
Slow down a sampling speed.
889
Cannot set On this module.
6.17
890
Settings can not be entered for channels on which no strain module is mounted.
–
891
Cannot be changed when Variable is turned ON.
Chapter 15
892
A DSP channel cannot be specified for this trigger type.
Chapter 15
893
Cannot be set if Operation is set to something other than KnockFlt.
Chapter 15
894
Cannot set this module when Operation is set to KnockFlt.
Chapter 15
895
Cannot be set when the trace is set to a frequency module.
Chapter 15
896
Cannot be set when the trace is set to a DSP channel.
Chapter 15
897
The capture window cannot be changed while the dual capture is in progress, and while the measuring 7.6
is in progress.
898
Cannot be set while the capture window is opened.
7.6
System Operation Errors (900 to 999)
Code
Message and Corrective Action
Section
901
Failed to backup setup data. Will initialize. Backup battery may be low.
7.10
902
System RAM failure. Maintenance service is required.
–
903
System ROM failure. Maintenance service is required.
–
906
Fan stopped. Power off immediately. Maintenance service is required.
–
907
Backup battery is flat. Maintenance service is required to replace the back-up battery.
3.4
908
Internal temperature is too high. Power off immediately. Maintenance service is required.
3.4
909
Illegal SUM value. Maintenance service is required.
3.4
910
RAM read/write error. Maintenance service is required.
–
911
Memory bus error. Maintenance service is required.
–
912
Communication driver error.
–
914
Time out occurs in Communication.
–
915
EEPROM read error. EEPROM may be damaged. Maintenance service is required.
–
916
EEPROM write error. EEPROM may be damaged. Maintenance service is required.
–
917
No module installed. Install the module.
3.3
918
Turn ON the internal hard disk motor.
13.2
919
The current module installation condition and the setup data are inconsistent.
–
Will initialize.
922
Cannot back up the acquisition memory. Memory will be initialized.
7.10
The battery for acquisition memory backup may be flat.
928
Battery for waveform memory is flat. Waveform memory was initialized.
7.10
929
MS bus error occurred.
–
930
Slot 1 EEPROM error.
–
931
Slot 2 EEPROM error.
–
932
Slot 3 EEPROM error.
–
933
Slot 4 EEPROM error.
–
934
Slot 5 EEPROM error.
–
935
Slot 6 EEPROM error.
–
936
Slot 7 EEPROM error.
–
937
Slot 8 EEPROM error.
–
938
Key protect is enabled. To release the protection, press the PROTECT key.
17.6
939
This firmware is for the DL750P. It cannot be used on the DL750.
17.6
940
The USB device’s power consumption exceeded the capacity of the USB hub.
19.8