![YOKOGAWA ScopeCorder DL750 Скачать руководство пользователя страница 292](http://html.mh-extra.com/html/yokogawa/scopecorder-dl750/scopecorder-dl750_user-manual_911347292.webp)
19-4
IM 701210-06E
19.5 Function
Vertical/Horizontal Axis Settings
Item
Specifications
Channel ON/OFF
Independently turn ON/OFF CH1 to CH16, DSP1 to DSP6, LOGIC A, LOGIC B, and EVENT.
ALL CH menu
Set all channels while displaying waveforms. Operation using the USB keyboard is possible.
Vertical axis expansion/
Expand or reduce the vertical axis for each channel.
reduction
Variable
Upper/Lower limit scaling when variable is ON.
Input filter
Set for each channel.
Vertical position setting
Waveforms can be moved vertically in the range of
±
5 div from the center of the waveform display
frame.
Linear scaling
Set AX+B mode or P1-P2 mode independently for CH1 to CH16.
Roll mode
The roll mode is enabled when the trigger mode is set to auto, auto-level, single, or log, and the time
axis setting is greater than or equal to 100 ms/div.
Analysis
Item
Specifications
Search & zoom function
Search for, then expand and display a portion of the displayed waveform.
Choose from the following two search methods.
Edge:
Counts the rising and falling edges and automatically searches an arbitrary edge
Auto scroll:
Automatically scrolls the zoom position.
History search function
Search for and display waveforms from the history memory that satisfy specified conditions.
Zone search:
Set an area on the screen, then extract and display only those waveforms that pass through the
area (Pass mode), or do not pass through the area (By Pass mode).
Cursor measurement
The following cursors are selectable.
Horizontal, Vertical, H&V (only during X-Y waveform display), Degree (only during T-Y waveform
display), and Marker
Automated measurement of waveform parameters
Capable of performing automated measurement of waveform parameters. Automated
measurement of waveform parameters within one period (P-P through Int2XY).
Up to 24 items can be displayed.
P-P, Amp, Max, Min, High, Low, Avg, Mid, Rms, Sdev, +OvrShoot, –OvrShoot, Rise, Fall, Freq,
Period, +Width, –Width, Duty, Pulse, Burst1, Burst2, AvgFreq, AvgPeriod, Int1TY, Int2TY, Int1XY,
Int2XY, Delay (between channels)
Statistical processing
Applicable items: Automated measured values of waveform parameters described above.
Statistics:
Max, Min, Avg, Sdv, and Cnt
Maximum number of cycles:
48000 cycles (when the number of parameters is 1)
Maximum total number of parameters: 48000 (total number of results)
Maximum measurement range:
10 MW
Normal statistical processing
Performs statistical processing on all acquired waveforms while acquiring waveforms.
Cycle statistical processing
Performs statistical processing per cycle (cycle statistical computation). Extracts a periodic waveform
(cycle) from the acquisition memory and automatically calculates waveform parameters per cycle.
Statistical processing of the history data
Performs automated measurement of waveform parameters on the waveform that is acquired using
the history memory function and performs statistical processing.
Statistical processing is performed from the oldest data.
Computation (standard)
Operators: +, –,
×
, /, binary computation, phase shift, and power spectrum
User-defined
Equations obtained by arbitrarily combining the following operators.
computation(optional)
ABS, SQRT, LOG, EXP, NEG, SIN, COS, TAN, ATAN, PH, DIF, DDIF, INTG, IINTG, BIN, P2, P3,
F1, F2, FV, PWHH, PWHL, PWLH, PWLL, PWXX, DUTYH, DUTYL, FILT1, FILT2, HLBT, MEAN,
LS-, PS-, PSD-, CS-, TF-, CH-, MAG, LOGMAG, PHASE, REAL, IMAG
Phase shift
Monitor waveforms by shifting the phase of CH1 to CH16. Computation is performed on the phase-
shifted result.
GO/NO-GO determination The following two types of GO/NO-GO determination are available
• Determination using zones on the screen
• Determination using the result of the automated measurement of waveform parameters
Specify an action for GO or NO-GO result. Possible actions are screen image data output,
waveform data storage, buzzer notification, and e-mail transmission.
1
1.
When the Ethernet interface option is installed