![YOKOGAWA ScopeCorder DL750 Скачать руководство пользователя страница 283](http://html.mh-extra.com/html/yokogawa/scopecorder-dl750/scopecorder-dl750_user-manual_911347283.webp)
18-12
IM 701210-06E
Code
Message and Corrective Action
Section
1605
Repeat trigger cannot be specified.
Chapter 9
Repeat trigger is valid only during chart recorder mode when the print style is waveform and shot
recording is not OFF.
1606
Cannot be changed during X-Y recording.
9.7
1607
Cannot use wave window trigger when the recorder mode is set to chart or X-Y.
Chapter 9
1608
Cannot be specified when the print style is Numeric.
9.6
1609
Cannot Set or Execute on the DL750.
Chapter 9
Other Errors (9999)
Code
Message and Corrective Action
Section
9999
This error No. is not defined.
–
Note
If servicing is required, initialize the instrument once for confirmation.
18.2 Messages and Corrective Actions