19-14
IM 701210-06E
19.11 General Specifications
Item
Specifications
Standard operating
Ambient temperature: 23
±
5
°
C
conditions
Ambient humidity: 55
±
10% RH
Within 1% of the rated error of the power supply voltage and frequency
After a 30-minute warm-up and after calibration
Recommended calibration 1 year
period
Warm-up time
At least 30 minutes
Storage temperature
–20
°
C to 60
°
C
Storage humidity
20% to 85% RH (no condensation)
Storage altitude
3000 m or less
Operating temperature
5
°
C to 40
°
C
range
Operating humidity range
20 to 85% RH (when not using the printer), 35 to 85% RH (when using the printer)
Operating altitude
2000 m or less
Rated supply voltage
100 to 120 VAC or 200 to 240 VAC (automatic switching)
Rated supply voltage
50/60 Hz
frequency
Permitted supply voltage
90-132 VAC/180-264 VAC
Permitted supply voltage
48 to 63 Hz
frequency range
Maximum power
Approx. 200 VAMAX (Maximum power when the printer is OFF and 16 channels are running is 135
consumption
VA (reference value))
Withstand voltage
1500 VAC for 1 minute across the power supply and earth
Insulation resistance
10 M
Ω
or more at 500 VDC across the power supply and ground
External dimensions
355 mm (W)
×
250 mm (H)
×
180 mm(D) (excluding the handle and other projections)
Weight
DL750:
Approx. 6.6 kg (only the DL750 with all options (/M3/C8/C10/P4) options, without the
chart paper)
Approx. 10.6 kg (DL750 + eight High-Speed 10 MS/s, 12-Bit Isolation Modules)
DL750P: Approx. 7.8 kg (only the DL750P with all options (/M3/C8/C10/P4) options, without the
chart paper)
Approx. 11.8 kg ( eight High-Speed 10 MS/s, 12-Bit Isolation Modules)
Module:
Approx. 300 g (High-Speed 10 MS/s, 12-Bit Isolation Module)
Instrument’s cooling
Forced air cooling. Exhaust on the left side panel and top.
method
Battery backup
Setup parameters and clock are backed up with the internal lithium battery
Battery backup
Approx. 5 years (at ambient temperature of 25
°
C)
Battery life
Fuse
Inside the power supply unit (cannot be replaced from the outside of the instrument)
Standard accessories for
Front panel protection cover
1
B8023EA
the DL750/DL750P
Soft case
1
B9946EB
Cover panels
8
B8023EN
Power cord
1
Printer roll paper (for the DL750)
3
B9988AE (10-m roll)
Printer roll paper (for the DL750P)
1
701966 (20 m roll)
AAA Alkaline batteries
4
A1070EB (LR6JECPT by Toshiba)
(for waveform memory backup)
Rubber hind feet
1
B9989EX(4 pieces (1 sheet))
User’s Manual
1
Operation Guide
1
Communication Interface User’s Manual
1
B8023YZ (CD-ROM)