![YOKOGAWA ScopeCorder DL750 Скачать руководство пользователя страница 360](http://html.mh-extra.com/html/yokogawa/scopecorder-dl750/scopecorder-dl750_user-manual_911347360.webp)
App-28
IM 701210-06E
SHARP Band-pass Filter Order Pass-band Width 15%
Center frequency
10%
11%
12%
13%
14%
15%
16%
17%
Order
155
110
89
73
62
52
49
41
Center frequency
18%
19%
20%
21%
22%
23%
24%
25%
Order
38
36
34
32
27
26
25
24
Center frequency
26%
27%
28%
29%
30%
Order
23
22
21
21
21
SHARP Band-pass Filter Order Pass-band Width 20%
Center frequency
12%
13%
14%
15%
16%
17%
18%
19%
Order
191
129
98
78
67
58
49
46
Center frequency
20%
21%
22%
23%
24%
25%
26%
27%
Order
40
38
36
31
29
28
27
26
Center frequency
28%
29%
30%
Order
25
24
20
Computation Delay
The group delay can be derived from the following equation. It is constant and depends
on the filter order.
Group delay = (Filter order – 1)/2
Unit: s/T
s
(where T
s
is the sampling frequency [s])
The computation delay can be derived from the following equation.
Computation delay = {4 + (filter order – 1)/2}
×
sampling period
However, if the sampling frequency exceeds 100 kHz, it is fixed to 100 kHz (10
µ
s).
It is also fixed to 100 kHz (10
µ
s) when in envelope mode.
Characteristics Examples
• SHARP Low-pass
-80
-60
-40
-20
0
0 f
0.1 f
0.2 f
0.3 f
0.4 f
0.5 f
5%
10%
20%
SHARP Low-pass Frequency Characteristics
dB
(f: sampling frequency [Hz])
Appendix 6 DSP Channel Computation (Optional)