11-52
IM 701210-06E
Explanation
By setting the upper/lower limits of the automated measurement of waveform
parameters, GO/NO-GO determination is performed on whether the measured value
enters the range or exits the range.
Setting the Determination Conditions: Setup
• Selecting the Target Waveform: Trace
Select the waveform from CH1 to CH16, DSP1 to DSP6 (optional), and Math
waveforms.
• Setting the Determination Mode
OFF:
Does not perform GO/NO-GO determination.
IN:
Go condition when the value enters the specified upper and lower limits.
OUT:
Go condition when the value exits from the specified upper and lower limits.
• Selecting the Parameter: Item
The waveform parameters that can be used are all the items of automated
measurement of waveform parameters (28 types) given in section 11.6. Up to 16
types of parameters can be determined simultaneously.
• Setting the Upper and Lower Limits of Parameters
The upper and lower limits vary depending on the parameters. They can be set in the
range –9.9999E+30 to 9.9999E+30.
• Setting the Logic
AND:
Executes the action when all parameter conditions (1 to 16 types) are met.
OR:
Executes the action when any one of the parameter conditions (1 to 16 types)
is met.
• Action Condition: ActCondition
Always:
Always executes the action.
Failure:
Execute the action when the GO condition is not met.
Success:
Execute the action when the GO condition is met.
When Always is selected, the operation specified by Action is executed each time the
trigger occurs. This function is useful such as when you wish to print the screen
image data to a printer each time the trigger occurs.
• Setting the Number of Actions: Sequence
Single:
Performs the action once.
Continue: Repeats the action up to acquisition count specified by ACQ Count. (If the
count is set to Infinite, the action is repeated until the waveform acquisition
is stopped.)
• Setting the Number of Determinations: ACQ Count
Set the number of waveform acquisitions.
Infinite:
Continues until the waveform acquisition is stopped using the Abort soft
key or START/STOP.
1 to 65536: Stops when the specified number of waveforms is acquired.
• GO/NO-GO Determination Using an External Input Signal: Remote
GO/NO-GO determination can be performed by synchronizing to an external signal
applied to the GO/NO-GO I/O terminal. The determination result can also be output
externally. Remote must be turned ON in order to perform GO/NO-GO determination
using an external signal.
11.9 GO/NO-GO Determination Using Measured Waveform Parameters