![YOKOGAWA ScopeCorder DL750 Скачать руководство пользователя страница 314](http://html.mh-extra.com/html/yokogawa/scopecorder-dl750/scopecorder-dl750_user-manual_911347314.webp)
19-26
IM 701210-06E
Universal (Voltage/Temp.) Module (701261) Specifications
Universal (Voltage/Temp.) Module (with AAF) (701262) Specifications
Item
Specifications
Standard operating
Temperature: 23
°
C
±
5
°
C
conditions
Humidity: 55%
±
10% RH
After a 30-minute warm-up and after calibration
Function
Temperature (thermocouple) or voltage measurement (switchable)
Effective measurement range [Voltage measurement]
20 div (display range: 10 div)
Number of input channels
2
Input coupling
TC, DC, AC, and GND
TC: Temperature (thermocouple) measurement
DC: Voltage measurement (DC coupling)
AC: Voltage measurement (AC coupling)
Voltage measurement
100 kHz
maximum sample rate
Temperature measurement 500 Hz
data update rate
Input format
Isolated unbalanced
Measurement range
[Voltage measurement] Voltage sensitivity:
5 mV/div to 20 V/div (1-2-5 steps)
[Temperature measurement] Thermocouple: K, E, J, T, L, U, N, R, S, B, W, and Au7Fe
Measurement
[Voltage measurement] Voltage sensitivity: 5 mV/div to 20 V/div (1-2-5 steps)
range/accuracy
1
Voltage accuracy:
±
(0.25% of 10 div)
[Temperature measurement]
2
Type
Measurement Range
Accuracy
K
–200 to 1300
°
C
±
(0.1% of r 1.5
°
C)
E
–200 to 800
°
C
Except
±
(0.2% of r 1.5
°
C) for –200
°
C to 0
°
C
J
–200 to 1100
°
C
T
–200 to 400
°
C
L
–200 to 900
°
C
U
–200 to 400
°
C
N
0 to 1300
°
C
R
0 to 1700
°
C
±
(0.1% of r 3
°
C)
S
0 to 1700
°
C
Except, 0 to 200
°
C:
±
8
°
C
200 to 800
°
C:
±
5
°
C
B
0 to 1800
°
C
±
(0.1% of r 2
°
C)
Except, 400 to 700
°
C:
±
8
°
C
Effective range is 400 to 1800
°
C
W
0 to 2300
°
C
±
(0.1% of r 3
°
C)
Au7Fe
3
0 to 300K
0 to 50K:
±
4K
50 to 300K:
±
2.5K
Frequency characteristics
1
(–3 dB point when sine wave of amplitude equivalent to
±
3 div is applied)
[Voltage measurement] DC to 40 kHz
[Temperature measurement] DC to 100 Hz
Maximum input voltage
4
(at a frequency of 1 kHz or less)
Both temperature and voltage input: 42 V (DC + ACpeak) (as a value that meets the safety
standard)
150 V (DC + ACpeak) (maximum allowable voltage, as a value that does not damage the instrument
when applied)
Maximum common mode voltage
5
(at a frequency of 1 kHz or less)
Both temperature and voltage input: 42 V (DC+ACpeak) (CAT I and CAT II, 30 Vrms)
DC offset
[Voltage measurement]:
±
5 div
maximum selectable range
Vertical resolution
[Voltage measurement] During voltage input: 2400 LSB/div
[Temperature measurement] When measuring temperature: 0.1
°
C
Lower –3dB point
[Voltage measurement]
0.5 Hz or less
when AC coupled
Input connector
Binding post
19.12 Module Specifications