Explanation
Limitations
Automated measurement of waveform parameters cannot be performed on the following
waveforms.
• Snapshot waveforms
• Accumulated waveforms other than the newest waveform.
• Sub waveforms measured using the dual capture function (version 3.10 or earlier.
However, if you load the measured data of Version 3.10 or earlier into a DL750/DL750P
of Version 4.01 or later, automated measurement of waveform parameters is possible.)
Measurement Parameters: Item Setup
You can select among the 28 types of measurement parameters shown below and delay
between channels. Up to 48000 combinations of parameters of all traces (CH1 to CH16
and Math1 to Math8) can be saved.
• Voltage-axis Parameters
P-P
Max
Min
High
Low
+Oshot
-Oshot
: Peak to Peak Value (MAX - MIN) [V]
: Maximum Voltage [V]
: Minimum Voltage [V]
: Root Mean Square Value
(1/ n )(
Σ
(xi)2)1/2 [V]
: Average Voltage (1/n)
Σ
xi [V]
: Standard Deviation
(1/n(
Σ
xi
2
–(
Σ
xi)
2
/n))
1/2
[V]
: Undershoot Value
(LOW - MIN)/(HIGH - LOW) x 100 [%]
: Overshoot Value
(MAX - HIGH)/(HIGH - LOW) x 100[%]
: High Level Voltage [V]
: Low Level Voltage [V]
: Amplitude (High-Low) [V]
: (MAX+MIN)/2
P-P
Max
Min
Rms
1
Avg
Sdev
(SDv)
2
–Ovr shoot
(
–
Ovr)
1
+Ovr shoot
(+Ovr)
1
High
Low
Amp
Mid
1. If Rms is ON on the channels (Math1 to Math8) on which power spectrum computation (PS or PSD) is
selected, the screen shows "Rms = overall value." For details on the power spectrum computation, see
sections 10.3 and 10.5. For details on the overall value, see page App-19.
2. ( ) shows the corresponding name at the measurement item setting screen.
• Time-axis Parameters
+ Width
- Width
Period
Distal Line (90%)
Mesial Line (50%)
Proximal Line (10%)
Rise
Fall
High (100%)
Low(0%)
: Rise Time [s]
: Fall Time [s]
: Frequency [Hz],1/PERIOD
: Period [s]
: Mean Frequency in
Measuring Range[Hz]
: Mean Period in
Measuring Range[s]
: Width (s) Greater than the Mesial Value
: Width (s) Smaller than the Mesial Value
: Duty Ratio + Width/Period x 100[%]
1. ( ) shows the corresponding name at the measurement
item setting screen.
Rise
Fall
Freq
Period
(Prod)
1
Avg Freq
(FR-A)
1
Avg Period
(PR-A)
1
+Width
(+Wd)
1
–Width
(
–
Wd)
1
Duty
Measurement
Range
Pulse : Pulse Count
Mesial Line (50%)
When Pulse=3
Set the measurement range
(T-Range) to match the pulse
width of the signal being measured.
(PlsN)
Distal Line (90%)
Proximal Line (10%)
11.6 Automated Measurement of Waveform Parameters