Autozero
The
Autozero
input can be used to set the signal of a channel to a zero
point, regardless of the current that is being measured.
Allow sufficient time after you performed the autozero and before
injection to ensure that the autozero command is completed.
Alternatively, an autozero may be performed during a run. In this case,
make sure that autozero occurs at a point in the chromatogram where
the response curve is expected to be reasonably flat, that is, 15 seconds
prior to the elution of a peak.
Divert
The
Divert
input can be used to control the stream-switching valve with
a third-party driver.
11.2.2
Relay Outputs
To use output contact closures, connect the outputs according to theirs
+
or
–
polarity. Connect the third wire of the I/O cable to the ground
below the polarity of the output to which it belongs.
Pump Off
The
Pump Off
output contact is used to send a signal to the pump or
other external device to stop the flow of mobile phase and turn off the
pump. This signal may be used to turn off the pump to prevent the
detector from being flooded with mobile phase if the gas pressure falls
below 50% of the required gas volume for more than 60 seconds.
To use the
Pump Off
contact closure, the HPLC system must be able to
accept an input from the detector that can be programmed to turn off
liquid flow. In some cases, a special cable is required.
CC1 and CC2
The
CC1
and
CC2
contact closures can be used to start an external
device, for example an autosampler, or can be used to change a valve
position. The two ports act in similar fashion to the
Pump Off
output.
TIP
If you use the Chromeleon software, relay output
CC1
corresponds
to the relay signal
CAD_Relay_1
(default name), and relay output
CC2
corresponds to the relay signal
CAD_Relay_2
(default name). The
Pump
Off
relay output is not available in the Chromeleon software.
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...