6
•
Operation
Page 152
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
To monitor the baseline signals:
1. Open the ePanel Set.
2. On the toolbar above the ePanel Set, click
Monitor Baseline
.
3. In the
Select Channels to Monitor
dialog box, select the signals that
you want to monitor.
4. To stop monitoring the baseline, click
Stop
on the toolbar.
TIP
Monitor baseline data is overwritten each time baseline monitoring
is started. If you want to save the data permanently, define the location
in the
Monitor Baseline Save Preferences
dialog box. For details, refer
to the
Help
.
Running Automated Sample Analysis
Automated sample analysis with the chromatography system involves
creating and running a sequence. In a sequence, you determine how the
injections of a sample are analyzed and in which order they are
processed. Each injection of a sequence is processed by using an
instrument method, which defines the property values and timed
control commands for each module. For details how to create an
instrument method using the Instrument Method Wizard, refer to the
Help
.
Creating the sequence
Chromeleon 7 supports two methods for creating a new sequence:
•
eWorkflows (preferred method)
eWorkflows provide a set of predefined templates and rules for
creating a new sequence.
TIP
If no eWorkflows are available, you can create a new eWorkflow as
described in the
Help
. As an alternative, use the Sequence Wizard to
create the sequence.
a) On the
Console
, click the
eWorkflows
Category Bar.
b) In the
Navigation Pane
, click the eWorkflow name.
c) In the Work Area, click the instrument name, and then click
Launch
.
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...