5
•
Installation
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
Page 71
1. Place the detector in the system as required by the system
configuration. For details, see
section 5.5.1 System Arrangement
with an UltiMate 3000 System, page 67
2. Verify that the power switch on the detector is set to OFF.
3. Connect the power cord to the power-inlet connector on the
detector.
NOTICE
Before connecting the detector to the power line, be sure that
no condensation is present in the device. Condensation in the device can
damage the electronics. If you suspect that condensation is present,
allow the device to warm up to room temperature. This may take
several hours. Wait until the condensation is completely gone before
proceeding.
AVIS
Assurez-vous de l’absence de condensation dans l’appareil avant
de brancher le détecteur à la ligne électrique. La présence de
condensation dans l’appareil peut endommager l’électronique. Si vous
suspectez la présence de condensation, laissez l’appareil se réchauffer à
température ambiante. Ce processus peut durer plusieurs heures.
Patientez jusqu’à disparition complète de la condensation avant de
poursuivre.
4. Connect the free end of the power cord to an appropriate power
source.
5. Connect the interface cables to the detector:
♦
For the USB cable, see
♦
For the I/O connection cable, see
TIP
For an overview of connections with an UltiMate 3000 system, see
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...