Charged Aerosol Detectors Corona Veo (RS) Operating Manual
6.5.2.2
Corona Veo RS Detectors
When
•
After installation or replacement of the nebulizer
•
If the nebulizer gas pressure has fallen outside the specified
pressure for the nebulizer
•
If you use the gas regulation modes (only available when operating
the detector from a chromatography data system)
Preparations
Make sure that the nebulizer is installed.
Follow these steps
1.
If the pump flow is turned on and the detector is in Run Mode
a) Turn off the pump flow. Wait several minutes for the detector
pressure to stabilize.
b) In the
Run Mode Menu
on the detector display, select the
Exit
soft key to stop the
Run Mode
and return to the
Main Menu
.
TIP
If you operate the detector from a chromatography data system,
disconnect the detector to unlock the display functions and access
the
Main Menu
.
2. In the
Main Menu
, press the
Gas
soft key to navigate to the
Gas
Regulator
screen and display the gas regulator parameters.
TIP
Write down the nebulizer gas pressure reading indicated as the
Pressure
parameter for later monitoring of the pressure.
3. On the
Gas Regulator
screen, select the
Setpoint
parameter and set
the pressure to the value that is defined on the gas pressure
certificate.
4. Select the
Save
soft key to save the new pressure setting. Press
Exit
to return to the
Main Menu
.
5. Wait until the gas pressure reading has stabilized and reached the
setpoint. On the
Diagnostics
screen, monitor the pressure reading of
the
Gas Pressure
parameter (see
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...