8
•
Troubleshooting
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
Page 229
8.5.4.3
Peak Shape and Response
Symptom
Possible Cause
Remedial Action
Inability to
autozero the
signal
Autozero on a peak or
void signal
The Autozero occurred on a relatively flat section of the
chromatogram.
Autozero on a very noisy
signal
Reduce the noise or increase the current range.
High background signal
on a very sensitive
current range
(background >10x gain)
•
Use a mobile phase that provides a lower background (lower
concentration of impurities). Observe the
Guidelines for the Detector in section 6.5.1, page 123
•
Increase the current range.
Peak broadening
Clogging of capillaries,
solvent line filter frits or
sample loop
•
Replace the capillaries.
•
Check the solvent line filter for permeability. Replace the filter
frit if necessary.
•
Replace the sample loop.
Column overloaded or
contaminated
Clean or replace the column.
Reduce the injection volume or injection mass.
Eluent has changed
Use freshly prepared solvent. Use HPLC-grade eluents only.
Extra-column volume
too large
•
If early peaks are broader than later eluting ones, check the
capillary I.D. and length, sample loop size, flow cell, for example.
•
Use a capillary with a smaller volume for the connection
between column compartment and detector.
Peak fronting
Sample or solvent is
stronger than eluent at
time of injection
Match sample-solvent strength to the eluent strength at time of
injection.
Peak tailing
Extra column volume
too large
Use short capillary connections with a suitable inner diameter.
Improper capillary
connections
Check capillary fittings for correct placement. Loosen and
retighten fittings. If you use Viper capillaries and fittings, refer to
the
Viper Installation and Operation Guide
.
Periodic baseline
fluctuation,
pulsation
Air trapped in the
system
Purge the system (refer to
Pump Operating Manual
).
Pressure fluctuations
from pump
Purge the pump; check general function (refer to
Pump Operating
Manual
).
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...