8
•
Troubleshooting
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
Page 231
8.5.4.5
Peak Area Precision
Symptom
Possible Cause
Remedial Action
Capillary problems Capillary connections
not installed properly or
not tight, dead volumes
in capillary connections
Properly install the capillary connections. Tighten the capillary
connections. Replace capillary.
Contamination or
carry-over
Autosampler carry-over
Refer to the
Autosampler Operating Manual
.
Contamination or carry-
over in the system
Flush the system using an appropriate solvent.
Environmental
conditions
Unstable and
decomposing sample
Use new sample. Change the conditions. Cool the sample in the
autosampler if possible.
Unstable environmental
conditions
Make sure that the temperature and air humidity are constant.
Use column thermostatting. Avoid draft.
Injection volume
variation
Operating issues with
the autosampler
Operating issues with the autosampler may include:
•
Air trapped in autosampler flow path
•
Autosampler draws air from vial
•
Draw speed too high
•
Gas content of sample too high or saturated
•
Injector needle clogged or needle tip deformed
•
Injection valve or other autosampler components leaking
•
Needle seat worn out
Refer to the
Autosampler Operating Manual
.
Pump problems
Operating issues with
the pump
Operating issues with the pump may include:
•
Piston seals in pump leaking
•
Air trapped in pump head
•
Baseline fluctuations
•
Pump pulsation or flow fluctuations
•
Irreproducible gradient
Refer to the
Pump Operating Manual
.
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...