5
•
Installation
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
Page 73
5.5.2.2
USB Port
This section describes how to connect the USB port of the detector to
•
A data system computer to operate the detector from a
chromatography software, such as the Chromeleon software
–and/or–
•
other system modules, such as UltiMate 3000 modules.
USB Connections
All USB connections require standard USB A-to-B type cable.
Figure 8: USB cable
The flat, rectangular end of the cable is the "A" connector. The smaller,
hexagonal end is the "B" connector.
TIPS
•
The USB standard limits the USB cable length to 3 meters. Each USB
device can be separated from the computer or next USB hub by no
more than 3 meters.
•
After
connecting the USB cables and
before
turning on the power to
a device for the first time, verify that the chromatography software
is installed on the computer. When the power is turned on, the
required USB drivers are automatically found and the Windows™
operating system can detect the device.
•
The USB port is USB 1.1 and USB 2.0 compatible.
•
To ensure trouble-free operation, use only the cables shipped with
the detector.
•
The detector has no USB hub on its rear panel. Thus, it can only be
connected to USB hubs of other modules in the UltiMate 3000
system or directly to the computer.
•
If the detector is to be operated in an UltiMate 3000 system
It is not possible to use the USB hub on the UltiMate 3000
autosampler for connection of the detector to the computer.
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...