Charged Aerosol Detectors Corona Veo (RS) Operating Manual
Page 117
To adjust a parameter, select the respective parameter in parenthesis.
The available parameters or pressure values are displayed. Select the
desired parameter and press the
Save
soft key to save your changes.
Parameter
Description
Options
Pressure
Displays the actual pressure of the gas supplied
to the nebulizer
Read-only
Setpoint
Specifies the nebulizer gas pressure as specified
on the certificate that was shipped with the
nebulizer
40 - 65.0 psig
± 0.1 psig (276 -
448 kPa, 2.8 -
4.5 bar)
Pressure Units
Defines the default unit setting for the gas
pressure
psi, bar, kPa
6.4.3.2
Date & Time Setup Screen
The screen provides functions for setting the current date and time and
resetting the gas filter change interval and next preventive maintenance
(PM) date.
Figure 25: Date & Time Setup screen (here: Corona Veo RS)
•
To change date or time, select the respective entry in parenthesis
and set the desired values. Press the
Save
soft key to save changes.
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...