6
•
Operation
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
Page 145
To select the gas regulation mode for your application, note the
following:
•
Analytical
mode
♦
The pressure for the gas regulator is set to the nebulizer gas
pressure setting as defined on the nebulizer certificate.
♦
For use with high pump flow rates within the pump flow range
allowed for the detector
♦
This mode is set as a default in Chromeleon.
•
MicroLC
mode
♦
The nebulizer gas pressure setting is automatically reduced by a
fixed fraction of the nebulizer pressure.
♦
This mode optimizes the nebulizer gas pressure for certain pump
flow rates and certain types of mobile phase compositions.
♦
For use with very low pump flow rates within the pump flow
range allowed for the detector
TIP
The resulting nebulizer gas pressure will be below the gas
pressure specified in the nebulizer certificate (for example, below
55 psi or 3.8 bar or 379 kPa).
6.7.6
Power Function Value
The signal of the detector has a wide dynamic range with over four
orders of magnitude. The nature of the response is inherently nonlinear.
The
Power Function Value
(PFV) applies a power transformation to the
data output of the detector. Power transformations can improve the
linearity of response over a given range of interest.
The transformation is applied to the data output in real time. The data
output is raised to the selected power.
As a standard, the
Power Function
is set to 1.00. For most applications
and compounds, this is a good universal setting.
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...