Troubleshooting
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
Page 233
8.5.4.7
Liquid Flow
Symptom
Possible Cause
Remedial Action
Flow fluctuations
Pump flow path
For operating issues with the pump, refer to the
Operating
Manual
for the pump.
No flow
Column temperature
too high – eluent
evaporating
Select a lower column temperature (refer to
Column
Compartment Operating Manual
).
Leak in the system
Find and eliminate the leak.
Pump flow path
For operating issues with the pump, refer to the
Operating
Manual
for the pump.
8.5.4.8
Gas Flow Ratio and Flooding
Symptom
Possible Cause
Remedial Action
High gas flow ratio Clogging in the detector
Contact Technical Support.
Liquid in the gas
exhaust
Flooding inside the
detector (pump flow
turned on while no gas
flow present)
Flooding can occur if the detector is turned off while liquid is still
flowing into the detector.
See
section 8.5.3 Resolving Flooding In the Detector, page 224
Low gas flow ratio
Gas leak
Check the detector display for the respective error code. See
section 8.4 Warning and Error Codes, page 214
8.5.4.9
Display
Symptom
Possible Cause
Remedial Action
No information on
the display
Brightness on the screen
not adjusted
Adjust the display brightness. See
Error in electronic
system
Contact Technical Support.
Fuse has blown
Replace the fuses. See
section 7.10, page 196
If the replacement fuses fail, there is a fault with the power
supply. Contact Technical Support.
Module not connected
to the mains
Connect the power cord.
Module power turned
off
Turn on the module with its main power switch.
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...