6
•
Operation
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
Page 135
6.5.4
Detector Equilibration
6.5.4.1
Turning On the Detector Gas Flow
Follow the instructions in this section to turn on the detector gas flow.
CAUTION—Escaping Gas
Gas can escape from open gas connections when the detector gas flow
and/or the gas supply is turned on. The escaping gas can pose a health
and safety risk.
Turn on the gas supply and detector gas flow only when the nebulizer is
installed, the components in the gas flow path are properly connected,
and the protective cover of the gas filter assembly is installed.
ATTENTION—Fuite de gaz
Du gaz peut s’échapper des raccordements de gaz ouverts lors du
démarrage du débit de gaz du détecteur et/ou de l’alimentation de gaz.
Cette fuite de gaz peut présenter un risque pour la santé et la sécurité.
Démarrez l’alimentation de gaz et le débit de gaz du détecteur
uniquement lorsque le nébuliseur est installé, lorsque les composants
dans le trajet d’écoulement de gaz sont correctement raccordés, et
lorsque le capot de protection du bloc de filtration de gaz est installé.
NOTICE—Detector Gas Flow and Pump Flow
Improper handling of detector gas flow and pump flow start and stop
can damage the detector. Observe the respective order of steps:
•
To stop pump flow and detector gas flow
Turn off the pump flow first, wait at least 5 minutes and only then
turn off the detector gas flow. During the waiting period, the drain
pump drains any remaining mobile phase to waste.
•
To start detector gas flow and pump flow
Turn on the detector gas flow first, wait at least 5 minutes and only
then turn on the pump flow.
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...