7
•
Maintenance and Service
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
Page 191
13. Turn on the detector with its main power switch. Observe the notes
in
14. On the
Date & Time Setup
screen of the detector display, press the
Reset Filter Chg Date
softkey to reset the gas filter replacement
interval for 12 months.
15. Turn on the gas flow to check for proper functioning of the gas
filters. See
section 6.5.4.1 Turning On the Detector Gas Flow,
page 135
If no warning or error occurs, the gas filter replacement is
completed.
Before starting an analysis, let the detector equilibrate and be sure
that it is ready for operation (see
section 6.5 Preparing the Detector
for Operation, page 121