Maintenance and Service
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
Page 197
Follow these steps
The fuse holder is located next to the main power switch.
Figure 44: Fuse holder
No. Description
1
2
3
Main power switch (on/off control)
Power-inlet connector
Fuse holder (cartridge)
1. Use the screwdriver to remove the fuse holder.
2. Replace the two fuses with new fuses of the specified type and
current rating. Always replace
both
fuses.
3. Reinstall the fuse holder.
4. Reconnect the power cord to the power source and to the detector.
5. Turn on the detector with the main power switch.
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...