Follow these steps
1. On the stream-switching valve, loosen the 2 screws that attach the
stator with the wrench. Loosen the screws in an alternating pattern,
by only half a turn at a time to avoid that the stator gets jammed.
Figure 41: Stream-switching valve (view from front)
2. Remove the stator from the valve body. To ensure that the sealing
surface of the stator cap is not damaged, place the stator on its
outer face.
3. Carefully remove the rotor seal with your hand or a small
screwdriver.
Figure 42: Disassembled valve
4. Inspect the rotor and stator surfaces for scratches:
♦
If there are no signs of scratches, clean all parts thoroughly. See
♦
If there are signs of scratches, replace the rotor and/or stator
and reassemble the valve. See
No.
Description
1
Stator
2
Attachment screws for
stator
No. Description
1
Stator
2
Rotor
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...