8
•
Troubleshooting
Page 220
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
8.5
Operating Issues
8.5.1
Resolving Leakage
When
The leak sensor is wet. The leak sensor reports leakage.
Liquid may have collected in the leak tray under the flow connections.
Parts required
Replacement part as required
Additional items required
Cloth or tissue
Preparations
When resolving leakage, observe the safety guidelines and general rules
for maintenance and service as presented in
chapter 7 Maintenance and
Follow these steps
1. Turn off the pump flow and then the gas flow:
a) Turn off the pump flow to the detector. Wait at least 5 minutes.
During the waiting period, the detector drains any remaining
mobile phase to waste.
b) Turn off the gas flow on the detector.
c) Wait until the system pressure is down to zero before you
continue.
d) Turn off the gas supply to the detector.
2. If you operate the detector from the chromatography data system,
disconnect the detector from the software.
3. Turn off the detector with the main power switch.
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...