5
•
Installation
Page 68
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
Positioning the Detector in the System Stack
The arrangement of the system modules depends on the application.
Observe the following guidelines:
•
Place the charged aerosol detector in a position so that the
connections between column outlet and nebulizer inlet are
minimized.
A short distance between column outlet and nebulizer inlet reduces
post-column band broadening effects and optimizes
chromatographic resolution.
•
In UltiMate 3000 systems
If the charged aerosol detector is the only detector in the
UltiMate 3000 system stack, follow these steps:
1. To position the charged aerosol detector, grasp the detector by
the sides (in team effort) and position the detector on the
workbench. Place the detector in the lowest position in the
system stack.
2. Set up the system stack above the charged aerosol detector (see
For cable connections in an UltiMate 3000 system with a Corona
Veo (RS) detector, see
•
If an optical detector is used in addition to the charged aerosol
detector, the system arrangement may vary.
For details on series or parallel flow configurations with the charged
aerosol detector, see
section 5.6.5.3, page 94
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...