6
•
Operation
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
Page 137
Follow these steps
1. On the
Evaporation Temperature
screen, set the evaporation
temperature. To optimize the evaporation temperature for your
application, observe the guidelines in
2. To deliver pump flow to the charged aerosol detector, follow these
steps:
a) Make sure that the pump flow rate limit is set to the maximum
flow rate specified for the charged aerosol detector.
b) Turn on the pump flow to the detector. Have the pump deliver
flow to the detector.
Mobile Phase Guidelines for the Detector
The following table provides an overview of typical flow rates for
certain column internal diameters:
Column Inner Diameter
Typical Flow Rate
4.6 mm
1.0 mL/min
3.0 mm
0.5 mL/min
2.0 mm
0.2 mL/min
1.0 mm
0.05 mL/min
3. Allow the detector to equilibrate until you observe a steady
baseline.
♦
Monitor the background signal of the detector. It will increase
initially and should then start to fall. Over a short period of time,
the signal should fall to a low level and then stabilize.
♦
The detector is ready for data acquisition when the baseline is
stable. The time required to obtain a stable baseline depends on
the application.
4. Perform an autozero of the detector baseline. In the
Run Mode
Menu
of the display, press the
Autozero
soft key.
The detector is now ready for sample analysis.
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...