6.4.2.3
Evaporation Temperature Screen
The
Evaporation Temperature
screen provides the means to set and
monitor the temperature of the evaporation tube. To adjust a
parameter, select the respective parameter in parenthesis.
If no errors or warnings are active, the
Status
is
Normal
.
Figure 22: Evaporation Temperature screen (here: Corona Veo RS)
Parameter
Description
Temperature
Displays the actual evaporation temperature (read-only)
TIP
The actual temperature reached depends on several factors,
including mobile phase composition, liquid flow rate, mobile phase
temperature, gas temperature, and ambient temperature.
Setpoint
Allows setting the evaporation temperature
The default evaporation temperature is set to 35 °C.
TIP
Before you start sample analysis, wait approximately 30 minutes
after you changed the temperature setpoint to ensure that the
evaporation tube comes to equilibrium for the set temperature.
The available temperature settings depend on the detector:
Detector
Temperature setpoints
Corona Veo:
Selectable values: 35 °C or 50 °C
Corona Veo RS:
Settable range: a5 °C to +100 °C, in
0.1 °C steps
Power
Turning on or off the temperature control for the evaporation tube
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...