Troubleshooting
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
Page 213
Self-test
code
Description
Remedial Action
8
An error occurred with the corona charger
current when the voltage of the corona
charger was turned off.
Contact Technical Support.
10
An error occurred with the reading of the
flow ratio transducer at zero pressure.
20
An error occurred with the reading of the
pressure transducer at zero pressure.
40
An error occurred with the reading of the
total flow transducer at zero pressure.
80
An error occurred with the ion trap supply
upon power-up.
•
Liquid is present in the ion trap, indicating a
flooding in the detector. Turn off the pump flow.
Check that the waste and drain lines are free
from loops, kinks or blockages that may cause
the liquid to flow back into the detector. Resolve
the flooding. See
•
Repeat the self-test. If the problem persists,
contact Technical Support.
100
An error occurred with the ion trap supply
when the ion trap voltage was turned off.
•
Liquid is present in the ion trap, indicating a
flooding in the detector. Turn off the pump flow.
Check that the waste and drain lines are free
from loops, kinks or blockages that may cause
the liquid to flow back into the detector. Resolve
the flooding. See
•
Repeat the self-test. If the problem persists,
contact Technical Support.
200
The detector failed to reach a steady-state
internal temperature of the electrometer
amplifier.
•
The detector requires additional warm-up time
before the self-test. Repeat the self-test.
•
If the error still appears, contact Technical
Support.
400
The clock in the detector is not set.
Contact Technical Support.
800
The clock is not running.
1000
The clock battery is low.
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...