Installation
Page 98
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
5.7
Turning On the Detector
NOTICE
Make sure that the pump flow to the detector is turned off. To
properly turn on the detector gas flow and
only then
turn on the pump
flow to the detector, follow the instructions in
AVIS
Assurez-vous que le débit de la pompe vers le détecteur est éteint.
Pour démarrer correctement le débit de gaz du détecteur, puis
uniquement après cela
, démarrer le débit de la pompe vers le détecteur,
suivez les instructions de la
TIP
Before turning on the power to a UltiMate 3000 system module for
the first time, verify that the chromatography software is installed on
the data system computer. When the power is turned on, the required
USB drivers are automatically found and the Windows™ operating
system can detect the device.
1. Turn on the detector with its main power switch.
TIP
The drain pump turns on automatically upon power-up.
2. Wait until the detector completes the following sequence of events
after power-up:
♦
Self-test diagnostics
The detector runs a series of internal tests. During these self-
diagnostics, all of the main components are checked. When
testing is complete and has been successful, the display will
show the start-up screen with the current detector firmware
version. Select
Continue
to proceed to the warm-up screen.
If an error is detected during the self-test, the detector is not
ready for analysis, and the status of the self-test and a dedicated
error code appear on the detector display. For further
information, see
chapter 8 Troubleshooting, page 207
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...