Maintenance and Service
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
Page 177
Preparations
1. Flush the detector with the detector gas flow being turned on. For
recommendations on flushing the detector, see
section 7.4.3
Flushing the Detector between Analyses, page 172
2. Turn off the pump flow and then the gas flow:
a) Turn off the pump flow to the detector. Wait at least 5 minutes.
During the waiting period, the detector drains any remaining
mobile phase to waste.
b) Turn off the gas flow on the detector.
c) Wait until the system pressure is down to zero before you
continue.
d) Turn off the gas supply to the detector.
3. Turn off the detector with the main power switch.
Follow these steps
1. To disconnect gas tubes from the gas ports: Push the connected gas
tubing into the fitting and at the same time push the locking ring of
the gas port with the multitool. Then pull out the tubing.
Figure 31: Disconnecting the gas exhaust tubing with the multitool
(example)
2. To re-install the gas tubes or install new gas tubes, follow the steps
in
section 5.5.4 Connecting the Gas Inlet and Exhaust Tubes,
page 79
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...