Guidelines for selecting the evaporation temperature
The following guidelines are recommended for selecting an optimal
evaporation temperature:
NOTICE
Do not use non-volatile additives, such as sodium phosphate.
Observe the
Mobile Phase Guidelines for the Detector in section 6.5.1,
page 123
AVIS
N’utilisez pas d’additifs non-volatiles, tels que le phosphate de
sodium. Observez les consignes pour la phase mobile de la
section 6.5.1,
page 123
•
In general, use the lowest evaporation temperature within the
specified temperature range that provides acceptable limits of
sensitivity.
•
As a starting point, use an evaporation temperature of 35 °C.
•
Higher evaporation temperature settings can be used when analytes
of interest have low volatility, relative to the background. Since the
volatility of background residue is typically unknown,
experimentation is required for optimizing this parameter for a given
method.
•
Perform screening runs at two evaporation temperature settings:
♦
Corona Veo RS detector
Use a minimum of two evaporation temperatures. Typically,
35 °C and 40 °C are appropriate if the temperature range is
unknown. If required, perform additional tests in a maximum of
5 °C increments.
♦
Corona Veo detector
Use the two evaporation temperatures (35 °C and 50 °C) to
perform screening runs if the required temperature setting is
unknown.
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...