8
•
Troubleshooting
Page 224
Charged Aerosol Detectors Corona Veo (RS) Operating Manual
8.5.3
Resolving Flooding In the Detector
When
The liquid sensor inside the detector detected a build-up of excess liquid
(flooding) inside the detector.
The following actions occur:
•
A critical error code appears on the detector display and beeping
starts to alert you. See
section 8.4.2 Error Codes, page 216.
TIP
If you operate the detector from a chromatography data system,
an error message will also be documented in the audit trail of the
software.
•
The detector exits the
Run Mode
. If you operate the detector from a
chromatography software, the running sequence (or batch) is
aborted.
•
If not already turned on, the drain pump is turned on automatically.
•
If the detector is connected to a LC pump using the
Pump Off
output
on the I/O connection terminal, the pump will stop the pump flow.
•
Corona Veo RS only:
The stream-switching valve will divert the liquid
flow away from the detector.
Follow these steps
1. Turn off the pump flow to the detector if not already automatically
done by the
Pump Off
I/O connection.
2. Leave the detector gas flow turned on to dry the excess liquid. If
there is only a small amount of liquid present inside the detector,
the liquid may evaporate when leaving on gas flow overnight.
TIP
The period for leaving the gas flow turned on depends on the
severity of the excess liquid that built up.
Содержание Corona Veo
Страница 4: ...Contacting Us Page 4 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 10: ...Contents Page 10 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 100: ...5 Installation Page 100 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 206: ...7 Maintenance and Service Page 206 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 234: ...8 Troubleshooting Page 234 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 256: ...11 Appendix Page 256 Charged Aerosol Detectors Corona Veo RS Operating Manual...
Страница 257: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 257 12 Index...
Страница 261: ...12 Index Charged Aerosol Detectors Corona Veo RS Operating Manual Page 261...
Страница 262: ...Thermo Fisher Scientific Inc 168 Third Avenue Waltham Massachusetts 02451 USA www thermofisher com...