Setting the Tape Speed (SPD)
Set the tape speed by pressing the SPD/BCM,
;
and
:
keys. The recording times for the
corresponding tape speeds are as follows:
Tape speed
1 (normal speed)
1/2
1/4
1/8
1/16
Recording time (170 m tape)
2 h
4 h
8 h
16 h
32 h
Setting the Input Range (RANGE) (Analog channel)
Manual
1
To select the M-RANGE (manual range setting) mode, press the RANGE key.
2
Press the
;
or
:
key to select the channel.
3
Press the
8
or
9
key to select one of the input ranges below:
The left and right arrows in the table below indicate the direction of change when the key
is pressed.
Note
For example, the 20 V range is the range where the bar meter scale is ±100% or 0 dB when
the input voltage is +20 Vpk, –20 Vpk, or ±20 Vpk. The same applies to other values.
Automatic (After setting to E-E or REC-PAUSE mode)
1
To select the A-RANGE (automatic range setting) mode, press the SHIFT key and then
the RANGE key.
2
The peak value of the input level for each channel is automatically detected and held.
The relationship between the input levels and setting ranges are as shown below.
3
Press the RANGE key once after the barmeter shakes, is set the input ranges for all
channels corresponding to the last peak values being held.
Note
8
key
9
key
RANGE display
Input range (Vpk)
→
20
±20
←
→
10
±10
←
→
5
±5
←
→
2
±2
←
→
1
±1
←
0.5
±0.5
5.5 to less
than 11 V
±10
2.2 to less
than 5.5 V
±5
1.1 to less
than 2.2 V
±2
0.55 to less
than 1.1 V
±1
Less than
0.55 V
±0.5
11 V
or more
±20
Input level
Input Range
(Vpk)
When input signals include unnecessary DC components, set the input DC offset
(described next) before setting the input range.