Page 96
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
4. STAT
IM 5000 Calibration
STAT
IM
5000/5000S/5000 G4 Service Guide
4. This calibration will take approximately 6 minutes.
5. Allow the self-calibration to complete.
6. The temperature within the chamber will rise to the sterilization temperature. Wait until
sterilization phase of the calibration cycle ends automatically.
7. The offset value in the upper left-hand corner of the display (CVO) may have changed to a new
offset value.
8. Press the STOP button when finished.
3. Switch off the unit and remove all calibration equipment.