3. Troubleshooting Cycle Faults
>Temperature offset
Screen display when in Temperature Offset sub menu.
This sub menu is for information only.
## = unit chamber thermocouple offset value.
This allows the chamber thermocouple offset value to be read without entering calibration mode.
No changes can be made to the offset value in this sub menu.
1. To return to the service menu, press the RUBBER & PLASTICS button.
2. To return to the operational menu, press the STOP key.
>Press. Atm.offset – (S class units oNlY)
Screen display when in Press. Atm.Offset sub menu.
This sub menu is for information only.
## = unit pressure sensor (transducer) offset value at atmospheric pressure.
This allows the pressure sensor atmospheric offset value to be read without entering calibration
mode.
No changes can be made to the offset value in this sub menu.
1. To return to the service menu, press the RUBBER & PLASTICS button.
2. To return to the operational menu, press the STOP key.
>Press. Ster.offset – (S class units oNlY)
Screen display when in Press. Ster. Offset sub menu.
This sub menu is for information only.
## = unit pressure sensor (transducer) offset value for sterilisation conditions.
This allows the pressure sensor sterilisation offset value to be read without entering calibration
mode.
No changes can be made to the offset value in this sub menu.
1. To return to the service menu, press the RUBBER & PLASTICS button.
2. To return to the operational menu, press the STOP key.
STAT
IM
5000/5000S/5000 G4 Service Guide
>Temperature Offset
##
>Press. Atm. Offset
##
>Press. Ster. Offset
##