3. Troubleshooting Cycle Faults
STAT
IM
5000/5000S/5000 G4 Service Guide
>voltage Setup
Screen display when in voltage setup sub menu.
This sub menu is for information only.
This allows the unit voltage value to be read.
No changes can be made to the value in this sub
menu.
1. To return to the service menu, press the RUBBER & PLASTICS button.
2. To return to the operational menu, press the STOP key.
>Repeater Mode
Screen display when in Repeater mode sub menu.
Repeater mode allows you to run a selected cycle continuously for bench testing. The cycle
selected will run, vent and then dry for 2 minutes before re-setting and repeating.
The default setting on all STAT
im
units is
OFF
. To activate repeater mode, proceed as follows:
1. The UNWRAPPED or WRAPPED keys will both switch the screen display to “
ON
” (or “
OFF
”
depending on the opening display).
2. The RUBBER & PLASTICS key will save the change and return the unit to the service menu
condition.
3. Pressing the STOP key will exit without saving and return the unit to the operational menu.
>RS232
>end of line CR/lF
>Serial Port Bitrate
>Printer uses º char
>Factory Default
Screen display when in factory default sub menu.
This function will reset the NVRAM (Non Volatile Random Access Memory) to the factory default
settings.
>Voltage setup
115V
>Repearer Mode
On
>Factory default
No
At this point in the service menu you will see a repeat
of the items in the user menu used for setting up data
recording devices such as printers and Data Logger.
For details on setting up these items refer to the user
manual or Chapter 12 of the service manual.