![SciCan STATIM 5000S Скачать руководство пользователя страница 42](http://html1.mh-extra.com/html/scican/statim-5000s/statim-5000s_service-manual_1217148042.webp)
Page 42
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
2. Tools, Maintenance Schedules, Procedures and Testing
STAT
IM
5000/5000S/5000 G4 Service Guide
using the STAT
IM G4 remote access function
Users can allow offsite technicians to remotely access the LCD touchscreens and web portals of
STAT
IM G4 units connected to a network.
For local network remote access, proceed as follows:
1. From the TOOLS page, click on the LOCAL CONTROL tab.
2. Click on the start button to start a local connection. It will open up a page that mirrors your
STAT
IM unit’s touchscreen so that it can be controlled remotely within your local network.
For remote access of a STAT
IM web portal or touchscreen from outside a local network,
proceed as follows:
1. Someone onsite with the unit or from within the network must provide access to an outside
user by generating a ‘token’ (or code).
• To generate a unique token using the web portal, go to the TOOLS page and click on
the REMOTE ACCESS tab.