STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
2. Tools, Maintenance Schedules, Procedures and Testing
Figure 8.
Temperature fitting cut-away
If you have the SciCan supplied
Omega HH81A
type temperature reference meter with the
1/16”
(1.6mm) temperature probe
(normally used with the Non S-class type calibration cassette) then
the compression fitting attached to the S class cassette is correct and does
NOT
have to be
removed.
If you have either the
Fluke 51
type temperature reference meter, or
Heise PTE1
type combined
temperature and pressure reference meter then the 1/16” (1.6mm) compression fitting (complete)
will need to be removed from the extension piece attached to the cassette and replaced with the
1/8” (3.175mm) compression fitting from the accessory kit, as both of these meters have 1/8”
(3.175mm) thermocouple probes.
NOTE: PTFE tape should be used when assembling these parts to ensure a steam tight seal.
To insert the relevant temperature probe (thermocouple) into either of the appropriate compression
fittings, loosen the clamp nut and insert the probe into the fitting as far as it will go. A slight
resistance will be felt as the probe passes through the conical seal. Tighten the clamp nut until a
steam/air tight seal is achieved.
DO NOT OVERTIGHTEN
.
STAT
IM
5000/5000S/5000 G4 Service Guide