13. Installation
STAT
IM
5000/5000S/5000 G4 Service Guide
Setting/editing Serial Number
NOTE:
This function is available on Rev. 7 controller boards with software revision SXXXR612 on
STAT
im
Classic, and software revision SXXXR700 on STATIM G4.
For STAT
im
Classic, follow these steps::
1. Power
ON
the STAT
im
while holding down the UNWRAPPED and WRAPPED keys.
2. Enter the service password to access the service menu.
3. The default password is entered by pressing the UNWRAPPED, WRAPPED, RUBBER &
PLASTICS and
STOP
keys in that order (a backdoor password can also be used by pressing
the UNWRAPPED, WRAPPED, UNWRAPPED, WRAPPED keys in this order).
4. Enter the Service Menu and select EDIT SERIAL NUMBER from the list.
5. Use the UNWRAPPED button to increase the selected digit and the WRAPPED button to
decrease the selected digit.
6. Use the RUBBER & PLASTIC button to select the next adjacent digit.
7. To save changes and return to the regular operating mode, press the
STOP
button.
For STAT
im
G4, follow these steps:
1. Access the Service Menu as described in Using the
LCD Service Menu on the STAT
im
G4 in Chapter 3
and scroll to SET SERIAL NUMBER.
2. Use the first pair of the arrow buttons to move from
one digit to the next. Use the second pair of arrow
buttons to change digit values.
3. To save changes and return to the regular operating
mode, press the
STOP
button
NOTE:
The serial number can be set also by using
a Barcode Scanner connected onto the RS232 port
(Settings: 96008N1, CR/LF at the end of string).