STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
2. Tools, Maintenance Schedules, Procedures and Testing
STAT
IM
5000/5000S/5000 G4 Service Guide
Figure 14.
Assembled male fitting, female fitting and cassette extension fitting
Attaching the pressure tube to the recommended pressure reference meter (s)
Important notes:
• Attaching the pressure tube to the reference meter is achieved using the second male Swagelok
b-qc4-s-2pm fitting supplied with the cassette.
• Attaching the tube to the connector is as described in the section above, however, the second
male fitting will need to be installed onto the pressure reference meter prior to attempting this.
• How the second male fitting is attached will depend on the type of reference meter used.
Assembling the Swagelok b-qc4-s-2pm ‘quick connect’ fitting to the Druck dpi
705r absolute pressure meter external 0 - 7 bar absolute pressure transducer
with ¼” NPT female thread.
NOTE: To complete this assembly you will need to locate the 1/8 NPT female to ¼ NPT male
adapter supplied with the S-class calibration cassette. (See Figure 15 for assembly) .