![SciCan STATIM 5000S Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/scican/statim-5000s/statim-5000s_service-manual_1217148054.webp)
Page 54
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
3. Troubleshooting Cycle Faults
9. Using the “Unwrapped” and “Wrapped” buttons, scroll down or up the service menu list until
the sub menu title
“Lock Cycles”
appears next to the “>” cursor.
10. Now press the “Rubber and Plastics” button to enter the
Locking
sub menu.
11. Upon entering this menu the display will show the following for a short time:
And then, it will display all of the available cycles (7 cycles).
12. Press the “Unwrapped” button and the unit will
“lock in”
the cycles selected at steps 2, 3 and
4, and
“lock out”
the unselected cycles. The display will show for a short time:
And then, it will display
only
the cycles selected (3 drying) at steps 2,3 and 4 plus the
drying cycle.
13. Please note that if the “Unwrapped” button is pressed again the cycles will be unlocked and so
on.
14. To save changes and return to the regular operating mode, press the
STOP
button. The
changes will be saved
automatically
.
15. To quit without making changes, power the STAT
im
OFF
.
16. For unlocking the cycles on a unit with locked cycles, follow steps 6 to 10. Upon entering
locking/unlocking mode the display will show:
And display only the selected cycles.
17. Press the “Unwrapped” button to unlock the cycles.
18. The display will now show the following for a short time:
And then, it will display all of the available cycles (7 cycles).
19. To save changes and return to the regular operating mode, press the STOP button. The
changes will be saved
automatically
.
20. To quit without making changes, power the STAT
im
OFF
.
>Conductivity Setup
Screen display when in Conductivity setup sub menu.
STAT
im
units monitor water quality using conductivity. The default settings are based on the
minimum level required to detect water in the unit (except units fitted with a float – see below)
and the maximum level required to protect the internal components from damage due to mineral
deposits (e.g. Calcium) on the internal surfaces.
STAT
IM
5000/5000S/5000 G4 Service Guide
Cycle UNLOCKED
Available cycles
Cycle UNLOCKED
Available cycles
CD=xx.xuS/NNN/y.yppm
L=LL.L H=HH.H G=G.GG
Cycle LOCKED
Selected cycles
Cycle LOCKED
Selected cycles