![SciCan STATIM 5000S Скачать руководство пользователя страница 57](http://html1.mh-extra.com/html/scican/statim-5000s/statim-5000s_service-manual_1217148057.webp)
Page 57
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
3. Troubleshooting Cycle Faults
STAT
IM
5000/5000S/5000 G4 Service Guide
The default setting on all STAT
im
units is
NO
. To clear the list, proceed as follows as follows:
1. The UNWRAPPED or WRAPPED keys will switch the screen display to “
YES
” or “
NO
”
depending on the opening display.
2. The RUBBER & PLASTICS key will save your selection and return the unit to the service menu
condition.
3. Pressing the STOP key will exit without changes and return the unit to the operational menu.
>Display last CF number (#)
Screen display when in Display last CF# sub menu.
This sub menu is for information only.
This function will allow you to observe the last fault condition of the unit for diagnostic purposes.
1. The display will show the CF number (## above) and the cycle number when the fault occurred
((######)).
2. Pressing the RUBBER & PLASTICS key will return the unit to the service menu.
3. Pressing the STOP key will return the unit to the operational menu.
>Devices Test oN/oFF
Screen display when in Devices Test ON/OFF sub menu.
This function on revision 7 STAT
im
units allows the technician to activate a number of components
for diagnostic purposes.
The list of components is as follows:
• Pump (water)
• Solenoid Valve (cassette must be disengaged)
• Compressor
• Yellow LED (cycle LED)
• Extra 1L (latched digital output – 7000 only - not used in 2000/5000)
• Extra 2L (latched digital output – 7000 only - not used in 2000/5000)
• Drawer relay (STAT
im
900 units only – not used on 2000/5000)
When in the service menu (and unit in standby), the default condition of these components is
OFF
.
Activating (and de-activating) components is achieved as follows:
1. Using the UNWRAPPED (down) or WRAPPED (up) buttons will scroll through the sub menu to
the next component in the list.
2. Pressing the RUBBER & PLASTICS button will activate (switch on) and deactivate (switch off)
the chosen component.
3. Pressing the STOP key will return the unit to the service menu.
>Display last CF#
## (######)
>Devices Test On/Off
Pump Off