Page 201
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
9. Solenoid Valve and Pressure Transducer
STAT
IM
5000/5000S/5000 G4 Service Guide
1. Pressure transducer
2. Power switch
3. Line cord (not shown)
4. Transducer connector
5. Compression nut
6. Coiled transducer tube
7. Solenoid valve inlet “T” fitting
8. Compression nut
9. Nut and washer
10. Transducer bracket
11. Screw and washer
12. Chassis
1
4
6
2
7 5
9
10
8
11
12
Figure 6
Removing the Pressure Transducer
CAuTIoN:
the pressure transducer, steam generator and other components may be hot even
if the unit has not been operating. Guard against burns.
Check the atmospheric pressure reading on the independent reference pressure meter AND on
the unit display.